Skip to navigation Skip to content
FAQs

Using a Li-Cor LI-200SZ Pyranometer on a Symphonie Logger

The pyranometer signal output is a uA current; almost all other analog sensors output a voltage output.

The scale factor to enter into the logger or SDR software is determined as follows:

scale factor = 119.21 / cal sheet number

Example: calibration sheet says 93.7 uA output per 1000 Watts per square meter...

scale = 119.21/93.7
scale = 1.27

...enter 1.27 as the scale factor (slope) for the sensor channel...

Offset is always 0 for a pyranometer.

Instructions for wiring the pyranometer to the Symphonie Logger is covered in the Symphonie user’s manual.

Do not use the resistor supplied with the pyranometer when connecting to Symphonie. The supplied resistor is for connecting to a 9200-PLUS or Wind Explorer logger only.

 

Download Calibration Reports

Our Calibration Report Retriever Tool is a comprehensive database of reports for the following products:

NRG Anemometers
110S Temperature Sensor
200M Wind Vane
200P Wind Vane
BP20 Barometric Pressure Sensor
BP60 + BP60C Barometric Pressure Sensor
HybridMC Anemometer + Vane
HybridXT Anemometer + Vane
PVT1 PV Temperature Sensor
T60 + T60C Temperature Sensor
R1 Pyranometer
RH5X Relative Humidity Sensor
Thies First Class Advanced Anemometer
WindSensor P2546-OPR Anemometer

Tech Support

Whether you are troubleshooting in the field or learning how to install a product, we are here for you.