LOGR-S | CONFIGURING A DIFFERENTIAL ANALOG SENSOR

To configure a differential analog sensor to the LOGR-S:

1. Navigate to Sensors > Analog Channels

2. Select Configure on desired channel

3. Within the Analog Channel Configuration page: 

   • Select Sensor Type from dropdown menu

   • Description will auto-populate – edit as desired

   • Enter slope and offset

   • Enter serial number

   • Enter height of sensor

   • Enter elevation angle

   • Enter azimuth angle

4. Click Done

Note: A differential sensor will occupy 2 channels

5. Click Save

6. Listen for an audible alert (two beeps) from LOGR-S confirming channel configuration

7. Confirm sensor is on correct channel and check real-time readings on the Status > Sensor Outputs page