Skip to navigation Skip to content

Hukseflux LP02 Pyranometer

An economical class C solar radiation sensor from Hukseflux.

  • Hukseflux LP02 Pyranometer

Configuration Options

This thermopile-based sensor measures global horizontal irradiance and has a spectrally flat response across the full solar spectrum. The LP02 generates a small voltage signal that is proportional to the solar radiation flux, expressed in W/m2.

The LP02 is compatible with the SymphoniePLUS®3 Data Logger and SymphoniePRO® Data Logger when equipped with a Symphonie® SCM Card (SCM #6646, used with SymphoniePLUS3 or older) or Pro Signal Conditioning Module (P-SCM #9128, used with SymphoniePRO).

This kit includes a calibration certificate and removable 5 m (16.4') cable.

  Hukseflux LP02-C Pyranometer (#11147)
Description  
Sensor type Thermopile solar radiation sensor; ISO 9060:2018 ‘Class C’ (Second Class) compliant
Applications
  • Solar resource assessment and monitoring
  • Meteorological studies
  • Environmental monitoring
Sensor range 0 W/m2 to 2000 W/m2
Spectral range 305 nm to 2800 nm
Instrument compatibility All NRG Symphonie series loggers with:
  • P-SCM (#9128) for SymphoniePRO
  • Thermopile SCM (#6646) for earlier models (logger measurement range 0 W/m2 to 1500W/m2)
Output signal  
Signal type Microvolt proportional to total solar radiation
Transfer function Included on calibration certificate; for use with NRG Symphonie see application note "Using Thermopile Pyranometers on Symphonie Loggers"
Accuracy Class C (Second Class) per ISO 9060:2018
Calibration Calibration sheet included with each sensor defines output in microvolts per Watt/m2; traceable to World Radiometric Reference (WRR)
Calibration uncertainty < 1.8% (k=2) as specified by Hukseflux
Output signal range -0.1 mV to + 50 mV
Drift Less than 1% per year
Power requirements  
Power Required Zero (passive sensor)
Installation  
Mounting Mounts to tower using NRG's Mounting Plate | Pyranometer, Gen I (14502)
Tools required
  • 5/32 inch (4 mm) Hex Key
  • 5/16 inch (8 mm) wrench or nut driver
  • #2 phillips screwdriver
Environmental  
Operating temperature range -40 °C to 80 °C (-40 °F to 176 °F)
Operating humidity range 0 to 100%
Physical  
Connections Bare wire leads from cable connect directly to logger
see also logger user's manual for wiring diagram
Cable length 5 m (16.4 feet)
Weight 0.30 kg (0.66 lbs)
Dimensions 85 mm X 60 mm (3.3 inches X 2.4 inches)
Materials  
Cable Removable 5 m signal cable with molded Binder style four pole M12 connector (4-conductor cable with shield; included with sensor)
Detector Thermopile
Enclosure IP67 weatherproof anodized aluminum enclosure and stainless steel hardware

Customer Service

+1 802-482-2255

Mon – Fri: 8:30am – 5:00pm Eastern Time (GMT-5)

Symphonie® SCM Card

Symphonie® SCM Card

Condition your sensor’s output signal so that a SymphoniePLUS3 Data Logger can read it by installing a Symphonie Signal Conditioning Module (SCM).

Mounting Boom | .52M (20.5"), Pyranometer, Plane of Array

Mounting Boom | .52M (20.5"), Pyranometer, Plane of Array

Mount your pyranometers and achieve any orientation easily and securely.

Mounting Platform | 2xGHI for 2.2m Tower

Mounting Platform | 2xGHI for 2.2m Tower

Mount your pyranometers on one of our GHI Mounting Booms to orient the sensor to collect Global Horizontal Irradiance data.